Replacing the IonSABRE II probe sample capillary
January 11, 2016, 715004599 Rev. B 169
8. Unscrew the finger-tight PEEK union from the UNF coupling.
9. Remove the ferrule from the capillary.
10. Remove the capillary from the UNF coupling.
11. Dispose of the capillary and ferrule in accordance with local environmental
regulations.
Installing the new capillary
Required materials
• Chemical-resistant, powder-free gloves
• Needle-nose pliers
•7-mm wrench
• Combined 2.5-mm Allen wrench and cone extraction tool
• Red PEEK tubing
•LC pump
• HPLC-grade (or better) 1:1 acetonitrile/water
• Capillary
• Sharp knife or PEEK tubing cutter
• Safety goggles
Warning: To avoid spreading contamination, dispose of the capillary
and ferrule in accordance with local environmental regulations. The
used components can be contaminated with biohazardous or toxic
materials.