EasyManua.ls Logo

Weidmüller UCV/AR - Page 21

Weidmüller UCV/AR
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
21
Input calibration procedure
When the
display shows
Action/Description ENT
ï
Put the instrument in setup mode and scroll through the main menu
CIPn
Press Å or ì
CIPy
Calibrate inputs? Start No
Connect the current source to the inputs (for connections see over page).
I=0
Set the current source to 0.00 mA Calibrate Skip
I=5
Set the current source to 5.00 mA Calibrate Skip
I=10
Set the current source to 10.00 mA Calibrate Skip
I=20
Set the current source to 20.00 mA Calibrate Skip
I=50
Set the current source to 50.00 mA Calibrate Skip
After the display stops flashing connect the voltage source (for connections see over page).
E=0
Set the Voltage source to 0.00 V Calibrate Skip
E=5
Set the Voltage source to 5.00 V Calibrate Skip
E=10
Set the Voltage source to 10.00 V Calibrate Skip
E=20
Set the Voltage source to 20.00 V Calibrate Skip
E=50
Set the Voltage source to 50.00 V Calibrate Skip
FnC=
Continue with the setup sequence.
Note: To discard changes to calibration and setup you can remove the power at this point.
Terminal Connections for output calibration
Calibration Stage Signal type Terminal
Analogue Current Output
mA output + 14
mA output 15
Analogue Voltage Output
V output + 14
V output 16

Related product manuals