7 Hygienic preparation
6. Rinse parts with clean water.
7. Dry the parts carefully with a soft cloth.
8. If necessary: fill the base of the humidifier with fresh water
(see " Filling the respiratory air humidifier", page 34).
9. Insert the humidifier insert into the top of the humidifier.
Part Cleaning Disinfection Sterilization
Base of the
humidifier
With warm water and washing-up liquid.
Recommendation: Clean the housing
components in the top rack of the
dishwasher every week (maximum 65°C).
If necessary: Descale (see "7.5.1 Descaling
the respiratory air humidifier", page 74)
Immersion disinfection
(Recommendation:
gigasept FF
®
) or boil out for
5minutes
Not permitted
Top of the
humidifier
Wipe with a damp cloth: use water or mild
soap; do not use microfiber cloths
Wipe-down disinfection
(Recommendation:
terralin
®
protect or perform
advanced Alcohol EP) or boil
out for 5 minutes
Humidifier
insert
With warm water and washing-up liquid.
Recommendation: Clean the humidifier
insert in the top rack of the dishwasher
every week (maximum 65°C).
If necessary: Descale (see "7.5.1 Descaling
the respiratory air humidifier", page 74)
Boil out for 5 minutes
Element
If necessary: Descale (see "7.5.1 Descaling
the respiratory air humidifier", page 74)
Immersion disinfection
(Recommendation:
gigasept FF
®
)
Spray disinfection
(Recommendation:
perform advanced)
or boil out for 5 minutes