4. Press the “LB/KG” button until “SOFt” is displayed. Press “LB/KG” one more time; the firmware
version will be shown (U X.X.X). (See below reference.)
5. Record the firmware version in section 4, Device Record.
Scale setup
You can customize the scale to best suit your needs using the Custom Modes.
The custom modes on the scale are Setup, Service, Factory, Calibration.
1. Press ‘Zero’, ‘Reweigh’ and ‘Power’ together to enter Custom mode.
2. Key in the correct PIN to access the mode.
Some menu items include editors. These can be interacted with using the following keys: (e.g. Show
‘0000’ as initial 4 digits PIN):
REWEIGH: Increment the selected character. If the current selected character is “9”, it shall
wrap to “0” when the “Reweigh” button is pressed.
RECALL: Decrement the selected character. If the current selected character is “0”, it shall wrap
to “9” when the “Recall” button is pressed.
LB/KG: Move to the next character/enter.
Keying through the menu with the LB/KG key will show the option, then the current setting for that
option. To change a setting select the current setting for the desired option and use the REWEIGH
key to increment or move to the next setting option, or the
RECALL key to decrement or move to the
previous setting option. Set the selected option and move to the next setting with the LB/KG key.
Setup Mode
Device will enter setup mode by entering PIN 9821.
The scale will cycle through these parameters by pressing LB/KG key. Some of these options may be
turned off or on through the Factory Mode.
SOFt displays the software version.
dAtE displays the software release date in US date format (MM.DD.YY).
SCALE displays the scale model.