Directions for Use Chapter 3 Patient Monitoring 35
If MAP is enabled, MAP results are displayed in the message window.
The measurement display persists for two minutes or until another NIBP cycle is
initiated. If an error is detected, an error tone sounds and an error code appears in
the SYS window.
Automatic NIBP Measurement
Automatic NIBP measurements repeat continuously at programmed intervals.
To set up an automatic NIBP measurement, do the following:
1. Attach the cuff to the patient’s arm.
2. Press to set the measurement interval.
The two dashes (--) in the message window indicate that automatic measurement is
turned off.
3. To set an interval, press or to cycle through the options, which include - -, ST,
and a range of intervals: 1, 3, 4, 5, 10, 15, 30, 45, 60, 90, 120, and 240 minutes.
Note
If the SpO
2
sensor is attached and generating valid pulse rate data, then the
displayed pulse rate is derived from the SpO
2
sensor reading.
Note
The interval is the time from the beginning of one measurement cycle to the
beginning of the next measurement cycle.
Note
The ST interval selection works differently from the other intervals. For
information on using these settings, please refer to “STAT Measurement” on
page 38.
SYS
DIA
SpO2 %
/min
mmHg
ºC
ºC
M
INTERVAL --