80 Disassembly procedure Welch Allyn VSM 300 Series
2. Remove the temperature-board assembly from the SureTemp Plus module as
follows:
a. Remove the four screws securing the inside panel to the module housing.
b. Remove the four screws securing the temperature board to the module housing.
c. Tilt the temperature board and rotate it clockwise to free the cable connector
from the connector slot (J5) in the housing; then lift the board out of the housing.
Note
For older monitors where the temperature cable is routed through a ferrite bead
on the rear chassis:
If the temperature cable is pulled through the slot and out of the ferrite bead, it
will be difficult to reroute the cable through the bead. To avoid this difficulty, ...
Separate the temperature module housing from the rear chassis just far enough
to enable you to release the temp cable from the ZIF connector.
Four screws securing the inside
panel to the module housing
Four screws securing the temperature
board to the module housing