3. Restore factory default settings:
• To restore radio settings to factory default values, touch Radio settings.
• To restore all current monitor settings to factory default values, touch All settings.
A confirmation dialog displays.
4. Touch OK.
The factory default settings are restored.
Save the monitor configuration to a drive
You can save the monitor configuration to a USB flash drive. You can use the saved configuration
to restore this monitor’s configuration or to copy this monitor’s configuration to other monitors.
Note Not all flash drives are supported.
1. Connect a flash drive to the USB port.
2. Go to the Service screens as described in “Access the Service screens.”
3. Touch the General tab.
4. Touch Save to USB.
A confirmation dialog displays.
5. Touch OK.
The monitor configuration is saved to the USB flash drive.
Load a monitor configuration
You can load a configuration from a USB flash drive to the monitor.
Note Not all flash drives are supported.
1. Connect a flash drive to the USB port.
2. Go to the Service screens as described in “Access the Service screens.”
3. Touch the General tab.
4. Touch Configure from USB.
A confirmation dialog displays.
5. Touch OK.
The configuration from the USB flash drive overwrites the configuration on the monitor.
Enter an asset tag
You can enter an alpha-numeric identifier in the data field to serve as an asset tag for monitor
identification.
1. Go to the Service screens as described in “Access the Service screens.”
2. Touch the General tab.
3. Touch and enter up to 20 characters.
4. Touch OK.
22
Service menu Welch Allyn Vital Signs Monitor 6000 Series™