Temperature messages
Message Possible cause Suggested action
Alarm
Connect temperature probe. No probe is connected Connect a temperature probe
and retry.
The probe is faulty Replace the temperature
probe.
The temperature module returned a connect
probe message
Connect a temperature probe
and try again. If a probe is
already connected, replace
the probe. If the problem
persists, replace the
temperature module
Insert correct color-coded
probe well.
The probe well is missing Insert a temperature probe
well.
Replace temperature probe. The probe is faulty Replace the temperature
probe.
Temperature not functional. Call
for service.
An internal error occurred Review the error log.
Replace the temperature
module.
The USB cable is disconnected Check the USB cable.
The battery is depleted or missing (Braun
ThermoScan PRO 4000 thermometer only)
Replace the batteries.
Retry temperature
measurement.
An internal error occurred Review the error log for the
specific error. Check the user
settings. Verify that the probe
well is properly seated in the
probe well housing. Verify that
the probe is connected to the
monitor.
Retry the temperature
measurement. If the problem
persists, replace the probe.
Temperature time limit
exceeded. Retry temperature
measurement.
The direct mode timed out Return the temperature probe
to the probe well and retry
measurement.
Information
Tissue contact lost. Lost tissue contact while attempting to
acquire temperature reading or acquired
reading was performed with limited tissue
contact
Touch OK to dismiss the
message. Start a new
temperature reading.
40
Troubleshooting Welch Allyn Vital Signs Monitor 6000 Series™