Battery power manager messages
Message Possible cause Suggested action
Alarm
Low battery 5 minutes or less
remaining. (High-priority alarm)
Battery power is extremely low Plug the monitor into AC
power. If not plugged in, the
monitor automatically powers
off.
Powering down. Call for
service.
The battery experienced an internal error Replace the battery.
Battery is absent or faulty. There is no battery in the monitor Insert a battery.
The battery is faulty Replace the battery.
Information
Low battery 30 minutes or less
remaining.
The battery power is low
Touch OK to dismiss or plug
the monitor into AC power.
Device is operating in battery
mode.
The AC power cord has been disconnected
Touch OK to dismiss or plug
the monitor into AC power.
Configuration Manager messages
Message Possible cause Suggested action
Alarm
Unable to load configuration;
using factory defaults.
A configuration load error occurred Restore factory defaults. If the
error persists, call for service.
Functional error. Call for
service.
A critical configuration load error occurred Call for service.
Data did not transfer. The monitor is configured to send data on
save without being networked
Change advanced setting to
disable Automatically
send on manual save.
Information
No connection for send. The monitor is not configured to the network Change advanced setting to
disable Automatically
send on manual save
control.
Configure the monitor to the
network.
Service manual Troubleshooting 45