Welch Allyn 6000 Series - When Replacing the Top Housing

Welch Allyn 6000 Series
130 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Connector Connects with Connector Connects with
J1 Temperature module J21 NIBP module
J3 Printer J24 Communications board
J5 Sp02 module
For models with the temperature option, plug the leftmost USB cable into J1 on the main
board. Plug the other end into the temperature module. Insert the USB cable into the retaining
clip mounted on the temperature housing.
For models with a printer, plug the second from the left USB connector into J3 on the main
board.
Plug the small end into the USB printer port while pulling the plastic latch away from the
connector. Once the connector is seated, push the plastic clip over the connector to secure it.
Verify that the Ethernet cable passes freely through the space between the temperature
housing and the printer board, exiting the case in the space between the top of the
temperature housing and beneath the top housing.
Plug the communications board’s power cable from the short end of the main harness into J49
on the main board, with the ferrite bead closest to the main board.
When replacing the top housing on the chassis, line up the slots with the ends of the horizontal
struts in the top housing to ensure proper seating.
When replacing the top housing
For models without a printer, install a door blank as follows:
1. Remove the paper backing from the gasket on the blanking panel.
2. Install the door blank into the chassis, pressing to secure the blank to the chassis.
68
Disassembly and repair Welch Allyn Vital Signs Monitor 6000 Series™