Remove the temperature module or housing
No. Item No. Item
10 Screw, plastite #4-20 X 0.500 pan head 54 Probe well seal
37 Right side panel 55 Horizontal struts
42 Front housing 56 Temperature mounting cover
44 Temperature blanking panel 60 HSG mounting clamp, bottom housing
45 Thermal shield 67 Left insert
53 Temperature housing (templess) 71 Temperature connection access cover
Note These instructions also apply to removing the empty temperature housing for
models without the temperature option.
Note Do not disconnect the USB cable from the temperature module unless replacing the
module. Disconnect the USB cable at the main board when removing the module.
1. Remove the probe well from the temperature module.
2. Remove the top housing as described in “Remove the top housing.”
Note The temperature module may be removed and replaced without
disconnecting the main harness, battery harnesses, USB cables, or
light-bar harness from the main board.
3. Disconnect the power button flex cable from J6 on the main board and remove the securing
tape.