4. Remove the right side panel.
5. Remove the two screws (10) that secure the HSG clamp (60) and the bottom of the temperature
module with the temperature mounting cover.
6. Remove the two screws (10) that secure the top of the temperature module.
7. Remove the temperature module.
Note If you are replacing only the temperature module without replacing
the front mounting cover, stop here. Install the replacement module
in the reverse order of the disassembly procedure.
8. Do one of these:
If your monitor has a temperature module, remove the temperature mounting cover:
a. Remove the screw that secures the temperature mounting cover to the front housing.
b. Remove the temperature mounting cover (56) from the front housing.
If your monitor does not have a temperature module, remove the temperature blanking
panel:
a. Remove the two screws that secure the blanking panel to the front housing.
b. Remove the blanking panel (44).
72
Disassembly and repair Welch Allyn Vital Signs Monitor 6000 Series™