Material no. Description Qty
200-2000IN Pressure meter, NETECH 1
407672 BP Test volume repair fixture 113670 1
Off-the shelf item NetGear router a/b/g or equivalent 1
660-0138-00 Cable, patch 5' RJ45, T568B 1
6000-30 Single tube blood pressure hose, 5 ft. 1
About the Welch Allyn Service Tool
Use the service tool, Gold edition, to complete the full suite of functional tests.
For information about the service tool, see the following:
For instructions on installing and using the service tool, see the "Welch Allyn Service Tool
Installation guide" (part number 103820).
For instructions on performing the functional verification, see the service tool’s help files on
the Welch Allyn Service Tool CD (part number 103521).
Tests performed by the service tool
The service tool tests the host device and installed options as listed in the following table.
Test Description NIBP Temperature SpO2 Host
POST Performs the power-on self test (POST)¹
Firmware version Checks the firmware version
Firmware upgrade Loads the latest firmware into the module
Leak Verifies leaks using 100 cc volume
AD noise Checks noise on the pressure channel
Calibration Calibrates pressure transducers
Accuracy (NIBP) Checks the accuracy of transducers
across pressure range
Dump Checks dump valves
Inflation Verifies the pneumatic pump
Valve control Verifies control of the system valve
Inflation linearity Verifies operation of linear inflation
control hardware