EasyManua.ls Logo

Welch Allyn 9600 - Specifications

Welch Allyn 9600
16 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SPECIFICATIONS
Power Requirements
Voltage ll5 VAC +/-10%, 60 Hz (220VAC
+/-10%, 50 Hz for European)
Current
Heating 130 mA
After Warm-Up 60 mA
Internal Fuse Rating 1.6A
Power Consumption
Heating 15.6W
After Warm-Up 7.2W
Mechanical
Main Unit
Length 150 mm (5.9")
Width 80 mm (3.15")
Height 80 mm (3.15")
Power Pack (approximate dimensions)
Length 83 mm (3.25")
Width 61 mm (2.4")
Height 76 mm (3.0")
Weight 20 lbs
Environment
Storage Temperature 32ºF - 122ºF (0ºC - 50ºC)
Operating Temperature Range 65ºF - 80ºF (183ºC - 26.7ºC)
Operating Humidity 20% to 90% RH. Non-Condensing
Operating Specifications
Controlled Temperature of 9600
Low Setting 96.4ºF
High Setting 106.0ºF
Temperature Tolerance +/-0.1ºF
READY Indicator Temperature
Tolerance +/-0.1ºF
Heating and Cooling Times
Time to Heat to Low Setting from
Room Temperature 6 min.
Time to Heat to High Setting from
Low Setting 4 min.
Time to Cool from High Setting to
Low Setting 20 min.
15