ABPM 6100 Hardware Manual
DIR 80012720 Ver. F 21 / 28
6. Specifications
This section provides the unit specifications for the ABPM 6100 ambulatory blood pressure
monitor.
Power Requirements:
Two “AA” alkaline batteries or high capacity rechargeable batteries (NimH).
Dimensions:
124 x 70 x 33 mm (approximate).
Weight:
270g (including batteries, approximate).
Environmental Operating Condition:
Atmospheric Pressure Range of 700 HPA to 1060 HPA
Protection against electric shock:
Internally powered, Type BF
Operating Ambients:
Temperature: 10°C (50°F) to 50°C (122°F).
Humidity: 20% to 95% RH non-condensing.
Storage Ambients:
Temperature: -20°C (-68°F) to 70°C (158°F).
Humidity: 15% to 95% RH non-condensing.
Altitude: -170 to 1,700 meters
Gauging Ranges:
Systolic Blood Pressure: 60 to 250 mmHg
Diastolic Blood Pressure: 25 to 200 mmHg
Maximum Inflation: 270 mmHg
Heart Rate: 40 to 200 bpm
Memory:
Up to 250 readings using Alkaline batteries
Up to 110 reading using rechargeable NimH
Method of Measurement:
Oscillometric with step deflation.
Accuracy:
± 3 mmHg
International Standards:
EN 60601-1-1:2001 ; Medical electrical equipment Part 1-1: General requirements for
safety. Collateral standard: Safety requirements for medical electrical systems
Specific requirements for the protection against hazards of ignition of flammable
anesthetic mixtures and to control the risk of fire (Fire prevention)
EN 60601-2-30:2000 ; Medical electrical equipment Part 2-30: Particular requirements
for the safety, including essential performance, of automatic cycling non-invasive blood
pressuring monitoring equipment
EN 60601-1-2:2001 ; Medical Electrical Equipment Part 1-2: General requirements for
safety. Collateral Standard: Electromagnetic compatibility requirements and tests
EN 1060-1:1996 ; Specification for Non-invasive sphygmomanometers Part 1.
General requirements
EN 1060-3:1997 ; Non-invasive sphygmomanometers Part 3: Supplementary
requirements for electro-mechanical blood pressure measuring systems
AAMI SP 10 ES1 Category C: 1992 (battery powered) Electronic or automated
sphygmomanometers
Calibration:
Minimally, once per year
Safety System:
Maximum Inflation pressure limited to 300 mmHg.
Auto safety release valve for power failure.
Maximum BP measurement time limited to less than 180 seconds.
Data Connector: