42 Programming Welch Allyn AED 10 Automatic External Defibrillator
Setting Up the Internal Log
The Welch Allyn AED 10 has two modes of transmitting the internal log to a PC. Both
methods utilize the infrared data port located on the front of the unit. To select which
method to use, the user must access the Log: Setup Menu.
To change transmit mode
1. Access the Main Menu. Select Log to display the Log Menu.
2. Press the Next soft key to highlight Set Up. Press the Enter soft key to display the
Log: Setup Menu.
3. Press the Next soft key to change the transmit mode.
If using the SmartLink program, select SmartLink.
If transferring the service log, select Service. (See “Working with the Internal Log”
on page 31 for more information.)
4. Highlight the Save box and press Enter to save the changes and return to the Log
Menu.
5. To leave the transmit mode as it was originally, highlight the Cancel box and press
Enter to return to the Log Menu.
Volume
Contrast
Date
Time
Log
Supervisor
Main Menu
Setup
Back
Clear
Smartlink
Log Setup
Cancel
Service
Setup
Cancel
Save
Smartlink
Save