User Manual Programming 45
Setting the CPR Timer
The CPR timer begins if a “No Shock Advised” decision is made to time the appropriate
length that CPR should be performed.
To set the CPR Timer
1. Access the Supervisor Menu screen. Highlight CPR and press Enter. Select CPR
Timer to display the CPR Timer Menu.
2. Press the Next soft key to scroll through CPR timer lengths.
3. Highlight Save by pressing Enter after the desired time has been selected. Press the
Enter soft key again to confirm the change and return to the CPR Menu.
4. To leave the time as it was originally, press Next when the Save box is highlighted to
select the Cancel box. Press the Enter soft key to return to the CPR Menu.
Setting the CPR Tempo
The CPR Tempo is the rate at which a tone plays when the CPR Timer is running.
To set the CPR Tempo
1. Access the Supervisor Menu screen. Highlight CPR and press Enter. Select CPR
Te m p o to display the CPR Tempo Menu.
2. Press the Next soft key to scroll through CPR tempo values.
Program Allows the user to update to the current version of Welch Allyn
AED 10 software via the infrared port.
Restore Defaults Reset the Welch Allyn AED 10 options to factory defaults.
Unit ID Set the Welch Allyn AED 10’s Unit ID.
Dept Set the Welch Allyn AED 10’s Department Name.
Change Code Set supervisor password numbers.
Bit Rate Allows the user to change the bit rate of the infrared port.
Utility Opens the Utility Menu
Note
Available CPR timer values are 15, 30, 60, 90, 120 (default), 150, and 180 seconds.
CPR Timer
120
Save
Cancel
CPR Timer
150
Save
Cancel
CPR Timer
150
Save
CancelCancel
CPR Tempo
100
Save
Cancel
CPR Tempo
105
Save
Cancel
CPR Tempo
105
Save
Cancel