66 Specifications Welch Allyn AED 10 Automatic External Defibrillator
Physical
Electrical Isolation
Data Management
Dimensions 8 1/4" x 6 7/8" x 2 3/4" (210 x 175 x 70 mm)
Weight 3.5 (1.6 kg) pounds with battery
Operating Temperature 32ºF to 122ºF (0ºC to 50ºC)
Storage Temperature
(without battery)
-22ºF to 158ºF (-30ºC to 70ºC)
Humidity Up to 95% (non-condensing)
Altitude Up to 15,000 feet (-150 to 4570 m)
Shock Mil Std 810E method 516.4, procedure 1 (30G, 6-9ms pulse, 1/2 sine each axis)
Vibration Mil Std 810E method 514.4, category 1
Water resistance IEC 60529 IPX4
Drop Test Mil Std 810F method 516.5 Procedure IV (1 meter drop onto any edge, corner, or surface)
Power Unit operates on internal battery only
External Electrical Connections No external devices are attached to the unit
Risk Current Category Internally powered equipment with defibrillator-proof BF type patient applied part (as
per definition of IEC 60601-1 standard)
Event Documentation Internal
Internal Memory Capacity 1MB: 250 4-sec ECG samples or 3000 time stamped events
Quick Report Treatment Summary, Event Log, Test Log
Communication RS-232 via Infrared (IR) port to PC or printer