User Manual Specifications 69
Electromagnetic Compatibility
Guidance and manufacturer’s declaration—electromagnetic emissions
(IEC 60601-1-2 Table 201)
The Welch Allyn AED 10 is intended for use in the electromagnetic environment specified
below. The customer or the user of the Welch Allyn AED 10 should assure that it is used
in such an environment.
Medical Electrical Equipment needs special precautions regarding EMC and needs to be
installed and put into service according to EMC information provided in this document.
Category Standard Level
Radiated Emissions EN55011 CISPR11 B
ESD EN61000-4-2 8kV air 6kV contact
Radiated Susceptibility EN61000-4-3 10 V/m (20 V/m EN 60601-2-4)
Magnetic Field Emissions MIL STD RE101 (AAMI DF2 4.3.18.1)
Magnetic Field Susceptibility MIL STD RS101 1 Gauss, 47 Hz-1.8 kHz
Emissions test Compliance Electromagnetic environment - guidance
RF emissions CISPR 11 Group 1 The Welch Allyn AED 10 uses RF energy only for
its internal function. Therefore, its RF emissions
are very low and are not likely to cause any
interference in nearby electronic equipment.
RF emissions CSPR 11 Class B
Harmonic Emission IEC 6100-3-2 Not applicable
Voltage fluctuations/
flicker emissions IEC 61000-3-3
Not applicable