Safety notes
WARNING
!
The AM 232 Manual Audiometer is designed to be used
with a hospital grade outlet. Injury to personnel or damage to equipment
can result when a three-prong or two-prong adapter is connected be-
tween the AM 232 Manual Audiometer power plug and an AC outlet or
extension cord. Additionally, the AM 232 Manual Audiometer is equipped
with a specific power transformer which should not be interchanged with
any other transformer or supply.
This symbol indicates the location of a service adjustment part and
is intended for service personnel only. The AM 232 Manual Audiometer is
a specifically calibrated audiometer and the periodic service and adjust-
ments for the instrument that may be required should be done only by an
authorized Welch Allyn service technician.
CAUTION
!
The AM 232 Manual Audiometer is designed to comply
with the EMC requirements according to IEC 60601-1-2.
Radio transmitting equipment, cellular phones, etc. shall not be used in
the close proximity of the device since this could influence the perfor-
mance of the device. Particular precaution must be considered during
use of strong emission sources such as High Frequency surgical equip-
ment and similar device so that e.g., the HF-cables are not routed on or
near the device. If in doubt, contact a qualified technician or your local
representative. Refer to the Electromagnetic Compatibility (EMC) Guide
on CD 482-638702.