Mains to Isolated RS232 (Models 210/220 only): 4000 VAC
A. Plug modified RS232 cable (all wires connected) into the RS232 connector at rear of Atlas unit.
B. Connect the modified AC cord (hot, neutral, and earth ground connected) to the back of the Atlas Monitor.
C. Connect the black lead banana plug to the return jack of the Hypot 3500 Dielectric tester.
D. Connect the red lead banana plug to the HV jack of the Hypot 3500 Dielectric tester.
E. Apply power to 3500D. Press [Set]
F. Set voltage to 1500 VAC with [up]/[down] buttons
G. Press [Set]
H. Set current limit to 3 mA with [up]/[down] buttons
I. Press [Set]
J. Set Timer to 60 seconds with [up]/[down] buttons
K. Press [Set]
L. Set Continuity off.
M. Scroll through settings with [Set] button.
N. When settings verified, press [Exit] button.
O. Connect the red clip lead from HV Jack of Hypot 3500D to modified AC connector cord with all lines connected
(Hot, Neutral, and Ground)
P. Connect Blk Clip lead to common connection of joined RS232 leads
Q. Position electrical leads at rear of test bench
R. Press green [Test] button. Wait for 60 seconds.
S. Record current: _______mA