Section 3 - Troubleshooting
SERVICE MANUAL 6200-43E REV. A
WELCH ALLYN ATLAS MONITOR 3
46.Get Service Mode menu.
47.Press <Select> to highlight CRT test pattern.
48.Press <BP START/CANCEL> to show test pattern.
Examine display.
49.Press any key to end display.
50.Press <Select> to highlight Printer test pattern (if
fitted).
51.Press <BP START/CANCEL> to start test pattern.
52.Printer should print test pattern.
53.Press any key to end display.
54.Examine printout.
55.Press <Select> to highlight LED test .
56.Press <AUTO> to turn on all LEDs .
57.Press <BP START/CANCEL> to show automatic test
pattern.
58.Watch for a while, look for glitches in pattern.
59.Press <Set> to go to manual mode and step through
individual segments if needed to observe a problem.
60.Press <Select> to highlight Button test.
61.Press <BP START/CANCEL> to start test.
62.Press every button on system, <BP START/CANCEL>
last.
62.1. Verify that buttons match up with their names,
and that all buttons are functional.
62.1.1. If names don’t match, indicates
memory corruption: Replace main
board
63.Press <Select> to highlight Display A/D channels
(three or four screens worth)
63.1. Write down all values for later review.
63.2. Press <Set> and write down all values for
each screen.
64.Press <Trend> to return to Idle screen.
65.Connect the BP port to the BP simulator.
66.Set the simulator for a normal reading (140/80,
100BPM, NSR).
67.Press <BP START/CANCEL>
67.1. System should start pump, display manom-
eter value in SYSTOLIC LED; this value
should track and be very close to pressure
BP test
3.1 Functional Test and Initial Diagnostic