EasyManua.ls Logo

Welch Allyn AudioScope - 2.5 Calibration Procedure: Single Level AudioScope

Welch Allyn AudioScope
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
_________________________________________________________________________
2.5 Calibration Procedure: Single Level AudioScope
Incoming bench checks: Prepare the AudioScope for calibration.
2.5.1 ___ Check the tightness of the two slotted screws near battery
positive spring
2.5.2 ___ Check the tightness of the charger jack
2.5.3 ___ Remove RTV from the potentiometers
2.5.4 ___ Check that the sound tube extends at least 1/16" to 1/8" into
the cavity behind the lens. If it does not, move it upwards. Use
fingers. Long nose pliers could bend or crush the sound tube
See Figure 2.5.4 below
Preparation of Sound Equipment:
2.5.5 ___ Set up your sound equipment as per “AudioScope II Sound
Equipment Calibration and Setup Procedure A01825"

Related product manuals