2.7 Calibration Procedure, Multi Level AudioScopes
23300 (MICROCONTROLLER type, serial no.9670000 and higher)
Incoming bench checks: Prepare the AudioScope 3 for calibration.
2.7.1 ___ Check the tightness of the two slotted screws near battery
positive spring.
2.7.2 ___ Check the tightness of the charger jack.
2.7.3 ___ Check that the sound tube extends at least 1/16" to 1/8" into
the cavity behind the lens. If it does not, move it upwards with
long nose pliers taking care not to bend or crush the sound
tube.
Preparation of Sound Equipment:
2.7.4 ___ Set up your sound equipment as per "AudioScope II Sound
Equipment Calibration and Setup Procedure A01825"
Calibration:
Materials Required: Audiometer Handle Test Specifications A00277
1 sheet (This is the process.)
Electronics Module Test Specification, A00985
1 sheet (These are the specifications.)
Note :A note appearing on the specification reads"Exact Measurements are not
required for 100% inspection". This note does not apply to sound settings.
Adjust the AudioScope according to the printed sound settings and their
tolerances as shown on the print.
Your Sequence: Original Step number from A00277: Modifications if
any:
2.7.5 ____ step 1 Start Button Check for smooth
operation.
2.7.6 ____ step 2 On/Off Switch
2.7.7 ____ step 3 L.E.D.s LED.s must be bright.
2.7.8 ____ step 4 Rattle
2.7.9 ____ step 7 Cover
2.7.10____ step 8 Pneumatic Seal Check seal before sound.