ERROR CODE DESCRIPTION
1
Unknown error
2
This can indicate one of the following problems:
After 2 seconds the pulse has not reached the falling level
Heart frequency has not reached its defined range
Variation of the heart frequency is too large
Less than 4 measurements were taken
The rising level was not reached within 2 seconds after the falling level
Pulse bandwidth is out of range
Movement artefact
Maximum number of beats encountered in a stage
3
This can indicate one of the following problems:
The heart frequency was out of range in more than 3 pressure stages
The difference between systole and diastole was too small
The systole is considerably bigger than the pump pressure
An oscillation value was smaller than the start or end value
The pulse bandwidth is smaller than at start or end point
The MAP is out of range
The systole is out of range
The diastole is too large
The number of stages is too large
The oscillation value is too large
The total time is too long
A function took too much time
4
The battery voltage is too low
5
No cuff was detected
6
Too many measurements were taken—OR—No memory was available in EEPROM
7
No measurements were taken—OR—The measurements were stopped
8
Hardware failure—OR—Error at zero alignment or pumping started unexpectedly
CUFF SIZE
ARM CIRCUMFERENCE
SOFT DISPOSABLE CUFFS
ARM CIRCUMFERENCE
REUSABLE SLEEVE CUFFS
ARM CIRCUMFERENCE
TRADITIONAL REUSABLE CUFFS
Pediatric Cuff
Part 5082-94-3
15.8–21.3 cm
NA
Part 6100-10
16.0–21.8 cm
Small Adult Cuff
Par
t 5082-95-3
20.0–27 cm
Par
t 101340
18–27 cm
Par
t 6100-11
21.1–26.6 cm
Adult Cuff
Par
t 5082-96-3
25.3–34.4 cm
Par
t 101341
25–35 cm
Par
t 6100-12
25.4–34.3 cm
Adult Plus Cuff NA
Par
t 101342
33–40 cm
Par
t 6100-13
27.0–42.0 cm
Large Adult Cuff
Par
t 5082-97-3
32.0–43.4 cm
Par
t 101343
39–46 cm
Par
t 6100-14
34.3–48.2 cm
UNDERSTANDING ERROR CODES
SIZING CHART
Correct cuff sizing is vital to the accuracy of monitoring data.
Find the circumference of the patient’s upper arm and refer to
this cuff sizing chart for proper cuff selection.
NOTE: The codes listed in the table are
the codes displayed by the CardioPerfect
software. They are different from the error
codes which can appear on the ABPM 6100
LCD display.