EasyManua.ls Logo

Welch Allyn CareTemp Touch Free - Take a Measurement

Welch Allyn CareTemp Touch Free
30 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Take a measurement
1. Press Measure to power on the thermometer.
When the device is ready, a single dash appears on the screen, and a beeps sounds.
2. Position the thermometer between 4 and 6 cm from the center of the patient's
forehead, and aim between the eyebrows.
Note If the eyebrow area is covered with hair, sweat, or dirt, clean
the area and wait 10 minutes to before taking a
measurement.
Note Hold the thermometer and the forehead steady during
measurement. Movement can impact the measurement.
3. Press Measure. Slowly move the device toward or farther from the forehead.
The arrow on the display flashes until the correct distance is achieved.
Note After pressing Measure, the thermometer allows 20
seconds to capture the measurement. If you press Measure
again before the 20 second time limit is up, the process
restarts. If the measurement is not completed in 20
seconds, the device displays Err . Press Measure to restart
the process.
4. When the correct distance is reached the LED bar will illuminate, three beeps sound,
and the measurement is displayed.
To repeat the process, return to step 2.
Note The complete measurement or Err message displays for 3
seconds. After 3 seconds, the thermometer returns to ready.
5. The thermometer automatically turns off after 10 seconds of inactivity.
8
Using the thermometer CareTemp™ Touch Free Thermometer