• Add tape to the edge of the LCD frame, where the harness passes through to
connect to the LCD, to prevent abrasion to the LCD harness.
• When adding the ferrite beads, align the beads with the marks on the LCD frame.
• Add tape to secure the harness to the LCD frame where the harness passes
through.
Remove the front housing
Before you begin, remove the following as described in this manual:
• Battery
• Rear housing
• CO2 module (extended chassis only)
• SpO2 module or blanking panel
• NIBP module
• Bottom housing
• Top housing
• Light bar
• Temperature module and mounting cover
• Main board
• LCD
Remove the remaining strut from the front housing.
When replacing the front housing
For models with the temperature option, replace the thermal shield (45).
98 Disassembly and repair Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™