Material no. Description Qty Component
SRC-MAX Nellcor portable oximetry tester 1 Nellcor SpO2
4500-30 Blood pressure hose, 5 ft. 1 NIBP
200-2000IN Calibrated pressure meter, NETECH, or equivalent 1 NIBP
407672 BP test volume repair fixture 113670, or equivalent
Includes test manifold with bulb and valve
1 NIBP
620216 Fitting "Y" 1/8 X 1/8 X 1/8 1 NIBP
01802-110 Tester, calibration, 9600 plus 3 SureTemp, Braun
06138-000 CAL-KEY, assembly, M690/692 1 SureTemp
Create a blood pressure Y-tube
The blood pressure Y-tube is a piece of custom test equipment that connects the device
to the test setup. The Y-tube is composed of a modified blood pressure hose and a Y-
type fitting. Follow these instructions to create a Y-tube.
1. Cut a 4500-30 blood pressure hose approximately 6 inches from the connector that
connects to the device.
2. Split the end of the dual-lumen hose to create two separate hoses. Make sure not to
puncture either hose.
3. Insert one end of the Y-type fitting into each end of the hose.
NIBP test setup
For the NIBP leak test, overpressure test, or accuracy check, connect the test equipment
shown below. Connect the manifold to the volume repair fixture as indicated by the
service tool or the test procedure.
Service manual Functional verification and calibration 103