No. Item No. Item
1 Test volume repair fixture with test manifold,
bulb, and valve
4 Device
2 Pressure meter 5 USB 2.0/5-pin type A to mini-B cable
3 Blood pressure Y-tube 6 PC
NIBP leak test (Silver edition)
The NIBP leak test is performed automatically using the service tool. The leak test
pressurizes the system with a start pressure (P
s
) of 250 mmHg ± 10 mmHg. After 15
seconds (T
t
) the end pressure (P
e
) is measured. The leak rate is calculated using the
formula L = (P
s
– P
e
)/ T
t
. The test fails if the leak rate exceeds 5 mmHg in 15 seconds.
Perform an NIBP leak test
1. Set up the test equipment.
2. Launch the service tool on the PC.
3. If the simplified interface is active, choose the Service option.
4. Log on with your user ID and password or with ADMIN as the user ID and leave the
password field blank.
5. Power on the device.
6. Select the device you want to test from the device list.
7. Click NIBP Sensor under the Device Information tab.
8. Click Leak Test in the NIBP Sensor pane on the right side of the window.
9. Follow the prompts until the test completes.
10. Click Close.
11. To view the results, open the active log file by selecting File > View Log File >
Active log file > Ok.
12. To record the results of your test, go to “Service record.”
104 Functional verification and calibration Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™