Note The Braun option does not require disassembly because it is external. The
Masimo SpHb and RRa options do not require disassembly because they
are software upgrades.
Option install procedures
This table lists the procedures required to install all available internal options. For step-by-
step instructions, see “Disassembly and repair.”
CAUTION Before installing an option, disconnect patients and power
down the monitor.
Option to install
Radio CO2 SpO2
SureTemp
Plus Printer Procedure
Remove the battery.
Remove the rear housing.
Remove the communications board.
Remove the power-supply cover.
Remove the CO2 module or blanking panel.
Remove the SpO2 module or blanking panel.
Remove the NIBP module.
Remove the bottom housing.
Remove the top housing.
Remove the printer door blank.
Remove the temperature housing.
Remove the temperature blanking panel.
Remove the main board.
Remove the LCD.
Install the temperature cover.
Install the temperature module.
Service manual Options, upgrades, and licenses 121