EasyManua.ls Logo

Welch Allyn Connex 6000 Series - Factory Defaults

Welch Allyn Connex 6000 Series
185 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Factory defaults
General alarm
Settings Default value
Display alarm limits Enabled
Alarm audio on Enabled
Volume Medium
Advanced
Allow user to disable alarms Enabled
Allow user to turn off general audio Enabled
Minimum alarm volume Low
Nurse call threshold Medium
Audio pause time 2 minutes
SpO2 alarm condition delay 10 seconds
SpHb alarm condition delay 10 seconds
etCO2 alarm condition delay 15 seconds
Respiration alarm condition delay 10 seconds
No breath detected alarm delay
Adult: 30 seconds
Pediatric: 20 seconds
Neonate: 15 seconds
156 Appendices Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™

Table of Contents

Related product manuals