Service menu
Access the Service screens
Note You cannot access the Service screens if sensors or physiological alarms
are active or if vital sign measurements are displayed.
1. From the Home tab, touch the Settings tab.
2. Touch the Advanced tab.
3. Enter 6345 as the access code and touch OK.
4. Touch the Service tab.
The General screen appears.
5. Perform service tasks by making selections or touching other tabs.
Note Service tasks and how to do them are detailed in this
section.
6. When you are done, touch Exit.
The Home tab appears.
General tab
Restore factory default settings
1. Go to the Service screens as described in “Access the Service screens.”
2. Touch the General tab.
3. Restore factory default settings:
• To restore radio settings to factory default values, touch Radio settings.
• To restore all current settings to factory default values, touch All settings.
A confirmation dialog appears.
4. Touch OK.
The factory default settings are restored.
If you selected Radio settings, the radio reboots, and the device remains powered
on.
If you selected All settings, the device reboots.
23