Save the device configuration to a drive
You can save the device configuration to a USB flash drive. You can use the saved
configuration to restore this device’s configuration or to copy this device’s configuration
to other devices.
Note Not all flash drives are supported.
1. Connect a flash drive to the USB port.
2. Go to the Service screens as described in “Access the Service screens.”
3. Touch the General tab.
4. Touch Save to USB.
A confirmation dialog appears.
5. Touch OK.
The device configuration is saved to the USB flash drive.
Load a device configuration
You can load a configuration from a USB flash drive to the device.
Note Not all flash drives are supported.
Note If your configuration includes radio parameters, make sure the radio is
enabled. The radio must be enabled before you can import radio
parameters.
Note If a device setting is not available in the configuration file, the setting
returns to the factory default. This might be the case when the
configuration file was saved from a device with a different firmware
version.
1. Connect a flash drive to the USB port.
2. Go to the Service screens as described in “Access the Service screens.”
3. Touch the General tab.
4. Touch Configure from USB.
A confirmation dialog appears.
5. Touch OK.
The configuration from the USB flash drive overwrites the configuration on the
device, and the device reboots.
Enter an asset tag
You can enter an alpha-numeric identifier in the data field to serve as an asset tag for
device identification.
1. Go to the Service screens as described in “Access the Service screens.”
2. Touch the General tab.
3. Touch
and enter up to 20 characters.
4. Touch OK.
24
Service menu Welch Allyn Connex
®
Vital Signs Monitor 6000 Series™