Service record
Date: Time:
Device name: Serial number:
Technician: Service tool version:
NIBP tests
Leak test
Specification Actual reading Pass Fail
Max: 5
Overpressure test
Mode Specification Actual reading
Adult mode: 280 . . . . 329
Neonate mode: 130 . . . .164
NIBP accuracy check
Target pressure ± 5 mmHg Pressure meter Service tool Specification Pass Fail
0 mmHg mmHg mmHg ± 1 mmHg
50 mmHg mmHg mmHg ± 3 mmHg
150 mmHg mmHg mmHg ± 3 mmHg
250 mmHg mmHg mmHg ± 3 mmHg
118
Functional verification and calibration overview Connex
®
Spot Monitor