Item
No.
Material
No. Description Qty
Required Welch
Allyn purchase
Optional Welch
Allyn purchase
(see Notes)
7 Doc-10 Nellcor SpO2 extension cable 1 X
8
(Option 1)
Varies APM (configurations with Accessory
Power Management)
1 X
8
(Option 2)
Varies Power supply (configurations with no
APM)
1 X
9 106270 Service test box 1 X
10 106275 Micro USB cable 1 X
11 N/A Pressure meter (must include at least
two decimal points and be accurate
to within ±0.5 mmHg)
1
12 407672 Test volume repair fixture (113670)
with test manifold, bulb, and valve
1 X
13 N/A Blood pressure Y-tube 1 See instructions for creating the Y-tube in
the Basic functional verification section.
14 01802-110 Tester, calibration, 9600 Plus,
SureTemp
3 X
15 06138-000 Cal-key, assembly, M690/692 1 X
Note
All items listed in this table are required to complete a full functional verification and calibration. If
Welch Allyn does not make an item available for purchase, you must provide it yourself.
Note
You may substitute items marked as an optional Welch Allyn purchase with comparable OEM
products.
Full functional verification check and calibration tools
The list of tools below is what is required to perform a full device functionality check and
calibration. The tools are used in conjunction with the Welch Allyn service tool, Gold
license, to perform a device calibration.
Material no. Description Qty Component
01802-110 Tester, calibration, 9600 Plus 3 Braun
ThermoScan
PRO 6000 &
SureTemp Plus
probes
407672 BP test volume repair fixture 113670 1 NIBP
Service manual Functional verification and calibration overview 123