Available options
The following options can be added to each model’s base configuration.
CAUTION Before installing any option, disconnect the patient from the
monitor and power down the device.
Options
Model 7100 7300 7400 7500
Masimo
X
1 2
Nonin
X
2
Nellcor OxiMax
X
1 2
SureTemp Plus
Braun ThermoScan PRO 6000
Wireless radio
X X
Bluetooth radio
X X X
1
This configuration comes with either Masimo or Nellcor.
2
This configuration comes with either Masimo, Nonin, or Nellcor.
Standard with this model. Software upgrade available
for this model. Hardware upgrade
available only on a case-by-case
basis.
X No software upgrade available for this
model. Hardware upgrade available only on
a case-by-case basis.
Install options
All internal option installations entail opening the device case and performing some
disassembly. Because this process requires disconnecting internal components, Welch
Allyn requires that the device undergo a full functional test after reassembly and before
placing the device back in service.
Before installing a new option, read information about removing the option in
“Disassembly and repair.” After familiarizing yourself with the process, follow the
instructions in the disassembly section.
Upgrading software with the Welch Allyn Service Tool
The Connex Spot Monitor and the Welch Allyn Service Tool (WAST) interact through
Ethernet over USB technology, enabling the monitor to function as a USB device with
both a network and an IP address. This connection separates logically into two distinct
interfaces: a generic USB serial connection and a Remote Network Driver Interface
128
Options and upgrades Connex
®
Spot Monitor