Settings Default value
Systolic: Upper limit Adult: 220 mmHg (29.3 kPa)
Pediatric: 145 mmHg (19.3 kPa)
Neonate: 100 mmHg (13.3 kPa)
Systolic lower limit enabled Yes
Systolic: Lower limit Adult: 75 mmHg (10.0 kPa)
Pediatric: 75 mmHg (10.0 kPa)
Neonate: 50 mmHg (6.7 kPa)
Diastolic upper limit enabled Yes
Diastolic: Upper limit Adult: 110 mmHg (14.7 kPa)
Pediatric: 100 mmHg (13.3 kPa)
Neonate: 70 mmHg (9.3 kPa)
Diastolic lower limit enabled Yes
Diastolic: Lower limit Adult: 35 mmHg (4.7 kPa)
Pediatric: 35 mmHg (4.7 kPa)
Neonate: 30 mmHg (4.0 kPa)
MAP upper limit enabled No
MAP: Upper limit Adult: 120 mmHg (16 kPa)
Pediatric: 110 mmHg (14.7 kPa)
Neonate: 80 mmHg (10.7 kPa)
MAP lower limit enabled No
MAP: Lower limit Adult: 50 mmHg (6. 7 kPa)
Pediatric: 50 mmHg (6.7 kPa)
Neonate: 35 mmHg (4.7 kPa)
Advanced
Default view sys_dia_map
Tube type 2 tubes
Unit of measure mmHg
Algorithm default SureBP
Cuff inflation target (step algorithm)
Adult 160 mmHg (21.3 kPa)
Pediatric 140 mmHg (18.7 kPa)
Neonate 90 mmHg (12.0 kPa)
Service manual Appendix 141