Symptom Possible cause Suggested action
Perform NIBP functional checks.
Perform NIBP calibration checks.
Replace the pump assembly.
Replace the main board.
The NIBP module malfunctioned. Update the software.
Perform NIBP functional checks.
Perform NIBP calibration checks.
Replace the pump assembly.
Replace the main board.
The main board malfunctioned. Update the software.
Perform NIBP functional checks.
Perform NIBP calibration checks.
Replace the pump assembly.
Replace the main board.
The single or dual lumen switch failed. Ensure that the cuff is the
properly sized.
Clean the cuff connections.
Replace the cuff and tubing.
Replace the main board.
The NIBP module needs calibration. Update the software.
Perform NIBP functional checks.
Perform NIBP calibration checks.
Replace the pump assembly.
Replace the main board.
Attempt to upgrade the software
failed.
The hardware or software between the Host and
the Ethernet module is incompatible.
Retry software update.
Replace the main board.
The monitor lost communication with the
network or a timing issue occurred during the
upgrade.
Retry software update.
Replace the main board.
156
Appendix Connex
®
Spot Monitor