Message Possible cause Suggested action Alarm priority
SpO2 rebooting. 041f00 There is an SpO2 hardware failure A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
SpO2 rebooting. 042000 There is an SpO2 MCU failure A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
SpO2 rebooting. 042100 There is an SpO2 watchdog failure A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
SpO2 not functional.
042200
There is an invalid SpO2 board
type
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
that the SpO2 module be replaced,
and if problem is still present replace
the monitor's main board.
Very low
SpO2 rebooting. 042300 There is an invalid SpO2 master
control state
A malfunction is detected. There are
two possible causes of these
failures. First, the power provided to
the board is out of specification. In
this case, the failure may be cleared
once the underlying cause is
removed. Second, the board has an
actual hardware failure and recovery
is not possible. It is recommended
Very low
Service manual Appendix 171