EasyManua.ls Logo

Welch Allyn Connex Spot Monitor - Specify Server Settings

Welch Allyn Connex Spot Monitor
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WPA-Enterprise and WPA2-Enterprise Specify the following settings and then click Update
radio.
Anonymous identity Enable this option to disable the User name field.
User name Enter the EAP identity (maximum of 32 characters). This
is disabled EAP type TTLS.
Password Enter the EAP password (maximum of 32 characters).
This is disabled EAP type TLS.
Enable server validation Enable or disable server validation. This is disabled EAP
type PEAP-TLS.
EAP type Select the authentication protocol (TLS, TTLS, PEAP-
MSCHAPv2, PEAP-GTC, PEAP-TLS, EAP-FAST).
Option Select more specific EAP settings (Inner EAP Setting:
Auto or PAP, PAC Provisioning: Auto or Manual).
Update radio Touch Update radio to activate all new radio
settings not selected previously.
Touch OK in the confirmation popup.
Note None of the changed radio settings
take effect until you touch Update radio.
4. Do one of the following:
To continue in Advanced settings, touch another tab.
To exit Advanced settings and return to the Home tab, touch Exit.
Specify server settings
1. In Advanced settings, touch the Network > Server tabs.
2. Select the method used to identify the IP address of the server with which the
device will communicate.
3. Specify settings.
Setting Action/Description
Network connectivity method Select the preferred method to connect to the server.
Data encryption Turn the security data encryption on or off, and enable or
disable the server authentication.
Server authentication Turn the server authentication on or off.
Host IP Enable the device to connect to a host or other server at
a fixed IP address. In the IP address entry field, enter the
IP address. In the Port entry field and enter the port
number. The range of entry is 0 to 65535. Select either a
Welch Allyn host or a direct EMR connection.
NRS Enable the device to connect to a Network Rendezvous
Service (NRS) at a fixed IP address. In the IP address
entry field, enter the IP address. In the Port entry field
and enter the port number. The range of entry is 0 to
65535.
DNS hostname Enable the device to connect to a Network Rendezvous
Service (NRS) by entering a hostname to be sent to a
Domain Name Server (DNS) to retrieve the NRS IP
address. In the Hostname field, enter the DNS name. In
the NRS Service Port entry field, enter the port number.
The range of entry is 0 to 65535.