EasyManua.ls Logo

Welch Allyn Connex Spot Monitor - Page 40

Welch Allyn Connex Spot Monitor
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Save service logs to a USB flash drive
This procedure enables you to save the last 14 active log files to a USB
flash drive.
Note The Welch Allyn Service Tool enables you to save only the 2 most recent
log files.
1. In Advanced settings, touch the Service > Log file tabs.
2. Connect a flash drive to the USB port.
Note Not all flash drives are supported. See "USB flash drive
requirements" in the Appendix for details.
3. Touch Save to USB and follow the onscreen prompts.
Copies of the service log files are saved to the drive.
Restore factory default settings
Note This process does not restore the monitor to a custom configuration
provided at delivery. Instead, it deletes all custom configuration data and
restores factory default settings.
1. In Advanced settings, touch the Service > System tabs.
2. Reset to defaults:
To restore radio settings to factory default values, select Radio settings only.
To restore all current settings to factory default values, select All settings.
3. Touch Reset and follow the onscreen prompts.
The factory default settings are restored.
If you selected Radio settings only, the radio reboots, and the device remains
powered on.
If you selected All settings, the device reboots.
Review firmware status and load firmware from the network
1. In Advanced settings, touch the Service > System tabs.
2. In the right pane, review the firmware status information.
Note If the firmware is up-to-date, take no further action.
3. To update the firmware, touch Load and follow the onscreen prompts.
Note You can also load new firmware by rebooting the device.
34
Advanced settings Connex
®
Spot Monitor

Table of Contents

Other manuals for Welch Allyn Connex Spot Monitor

Related product manuals