EasyManua.ls Logo

Welch Allyn Connex Spot Monitor - When Replacing the SpO2 Insert

Welch Allyn Connex Spot Monitor
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reassembly notes
SpO2 insert options are Masimo, Nellcor, Nonin, or none.
Kit items:
105926, Masimo insert only
105927, Nellcor insert only
105929, Nonin insert only
105928, No SpO2 insert only
Note Reassembly is not the reverse of disassembly, but it is the same for each
insert.
When replacing the SpO2 insert
1. First complete the following steps:
Attach the power harness to the SpO2 insert. (See "Attach the power harness to the
insert.")
Install any new SpO2 board, if applicable. (See the reassembly notes in "Remove the
SpO2 board." )
Attach the power harness to the insert
Align the power harness with the insert and secure it with 2 T-10 Torx screws (see
the Torque Value Table).
Service manual Disassembly and repair 61

Table of Contents

Other manuals for Welch Allyn Connex Spot Monitor

Related product manuals