CAUTION Sterilizing the monitor could damage the device.
Note Disinfect according to your facility's protocols and standards or local
regulations.
Note See separate instructions for cleaning accessories.
Prepare to clean the equipment
CAUTION Some cleaning agents are not appropriate for all components of
the device. Use only approved cleaning agents. Using unapproved cleaning
agents may cause damage to components.
CAUTION Do not use bleach solutions of any kind when cleaning metal
electrical contacts. They will damage the device.
CAUTION When cleaning the device, avoid using cloths or solutions that
include quaternary ammonium compounds (ammonium chlorides) or
glutaraldehyde-based disinfectants.
Select an approved cleaning agent from the following tables.
Section 1. Approved for the Connex Vital Signs Monitor except the Braun
ThermoScan PRO dock
Cleaning agent
Additional information
Clorox HP
Dispatch Hospital Cleaner Disinfectant Towels with
Bleach (Clorox Healthcare®)
CleanCide Wipes (Wexford Labs)
70 percent isopropyl alcohol solution Applied to a clean cloth
10 percent bleach solution (.5% - 1% sodium hypochlorite) applied to a clean cloth
Section 2. Approved for the Connex Integrated Wall System except the Braun
ThermoScan PRO dock
Cleaning agent
Additional information
10 percent bleach solution (.5% - 1% sodium hypochlorite) applied to a clean cloth
70 percent isopropyl alcohol solution Applied to a clean cloth
Section 3. Approved for the Braun ThermoScan PRO dock
Cleaning agent
Additional information
Clorox HP
Directions for use Maintenance and service 185