Cleaning agent Additional information
70 percent isopropyl alcohol solution Applied to a clean cloth
Clean the accessories
1. Wipe the NIBP hose and any reusable cuffs with a damp cloth moistened in a mild
detergent solution.
2. Wipe the temperature probe with a cloth dampened with alcohol, warm water, or an
appropriately diluted, nonstaining disinfectant solution.
3. Clean the pulse oximetry sensors with a cloth dampened with 70 percent isopropyl
alcohol or 10 percent chlorine bleach solution.
4. Clean the RRa cable and the SpO2/RRa dual cable with a cloth dampened with 70
percent isopropyl alcohol.
5. Clean the EarlySense sensor by gently wiping with a soft, slightly damp cloth or
wipes containing antiseptic substances (alcohols, Chlorohexidine, and bleach
material up to 5% concentration). Ensure that the sensor is dry before re-use. Avoid
excessive liquids.
6. (Connex IWS only) To clean Welch Allyn 3.5V instrument heads attached to the
physical assessment handles, follow the instructions provided in their directions for
use.
CAUTION Never immerse any monitor accessories.