Message Possible cause Suggested action
Remove the sensor from the
patient and reapply.
SpO2 sensor expires in….
Note This message only appears
on devices configured with SpHb.
The SpO2 sensor will expire soon Replace the SpO2 sensor.
Temperature messages
Message Possible cause Suggested action
Connect temperature probe. No probe is connected Connect a temperature probe and
retry.
The probe is faulty Replace the temperature probe.
The temperature module returned a connect
probe message
Connect a temperature probe and
retry. If a probe is already
connected, replace the probe.
Insert correct color-coded probe
well.
The probe well is missing Insert a temperature probe well.
Replace temperature probe. The probe is faulty Replace the temperature probe.
Temperature not functional. Call
for service.
A module error occurred Call for service.
Temperature time limit exceeded.
Retry temperature measurement.
The direct mode timed out Remove the probe from the
measurement site.
Tissue contact lost. The probe has lost contact with the patient's
tissue
Touch OK to dismiss the
message. Return the probe to the
probe well and retry patient
temperature measurement.
Ensure that the probe makes
proper contact with the patient's
tissue.
Retry temperature measurement.
Note This message often
accompanies other temperature
messages.
A probe heater or data error occurred Retry the temperature
measurement. If the problem
persists, replace the probe.
User settings require adjustment Adjust the user settings and
retry.
The ambient temperature is out of range Operate the monitor within the
specified temperature range.
Retry patient temperature
measurement.
The SureTemp temperature module is faulty Call for service.