Specifications
Physical specifications
Protection classifications, all monitor configurations
Characteristic Specification
Electrical rating 100 – 240 V AC, 50 – 60 Hz, 0.8 –1.5 A
Duty cycle Continuous operation
Duty cycle – physical assessment handles 2 minutes on, 10 minutes off
Type of protection against electric shock Class I equipment (protectively earthed) with double insulation
Degree of protection against electric
shock, for parts applied to patients
Type BF defibrillator proof
IEC EN 60601-1
Recovery time following defibrillator
discharge
Less than or equal to 10 seconds
Flammable anesthetics
WARNING Not suitable for use with flammable
anesthetics.
Degree of protection provided by the
enclosure with respect to harmful ingress
of liquids
IPX1 (Vital Signs Monitor)
IPX0 (Integrated Wall System)
Vital Signs Monitor 6000 Series
Height Standard chassis: 10 in. (25.4 cm)
Extended chassis: 10 in. (25.4 cm)
Width Standard chassis: 11.4 in. (28.96 cm)
Extended chassis: 11.4 in. (28.96 cm)
Depth Standard chassis: 6 in. (15.3 cm)
Extended chassis: 7.5 in. (19.1 cm)