Note If patient data was saved during the most recent power
down and the device was in the Continuous Monitoring
profile, a Confirm patient and location dialog appears with
the following options:
Same patient and location
Same patient, new location
New patient
Note If "Allow profile change" is enabled in Advanced settings and
continuous sensors are attached to the device at startup, the
device powers up in the Continuous Monitoring profile,
overriding other default profile selections.
2. If the Confirm patient and location dialog appears, select the desire option and touch
OK.
If a system error is detected, the monitor becomes inactive until you press or until the
monitor shuts down automatically. The monitor displays a system fault message that
contains a wrench icon and a system fault code to aid service and engineers in
diagnosing the problem.
Set the date and time
1. Touch the Settings tab.
2. Touch the Device tab.
3. Touch the Date/Time vertical tab.
4. To change the date and time values: Touch the up and down arrow keys or touch
and enter a value.
Repeat for each value you want to change.
Note The date and time stamps on saved patient measurements
will adjust in response to new date and time settings.
Power down the monitor
This power-down method, which places the monitor into Standby mode, ensures that
patient measurements are retained in the monitor memory for a maximum of 24 hours.
These saved measurements are available for recall or sending electronically to the
network. This method also ensures that any configuration settings you have changed
and saved will be maintained at the next startup.
1.
Touch the Settings tab.
2. Touch the Device tab.
3. Touch Power down.
If there is a software update available, a system message asks if you want to
upgrade the software.
4. If you want to upgrade the software, touch OK.
The monitor continues to charge the battery when in Standby mode.
Directions for use Startup 57