30 Chapter 3 Reviewing the System Settings Welch Allyn CP 100 Electrocardiograph
Transferring a Configuration to Another
Electrocardiograph
You can transfer your configuration from one CP 100 electrocardiograph to another.
To Transfer a Configuration
1. Insert a storage device (SD memory card or removable USB device) into an
electrocardiograph that is configured as desired.
2. Press the Menu key .
3. Choose System Settings > Device Info > Manage Settings > Export Files >
Export Configuration Files.
The files are copied to the storage device.
4. Remove the storage device, and insert it into another CP 100 electrocardiograph.
5. Choose System Settings > Device Info > Manage Settings > Import Files >
Import Configuration Files.
The files are copied to the electrocardiograph, and then it reboots.
Figure 20. Configuration File Transfer
Configuration file
or