Service Manual CP 100 and CP 200 Electrocardiograph 59
5. Unplug the two battery connectors. See
Figure 49.
You may need to pull hard. Use pliers
gently if necessary.
6. Connect a new battery by matching
connector sizes.
Do not force connectors to mismatch.
7. Place the battery into the battery
compartment. Replace the battery door
and screws. Turn the electrocardiograph
right-side-up.
It turns on automatically and displays a
prompt to enter date and time.
8. Reset the unit after you replace the battery.
See Battery Accuracy Test (page 31) for
directions.
9. Enter the date and time.
The electrocardiograph is ready to use.
10. Discard the old battery appropriately.
Figure 49. Battery Connectors
Procedure
Battery
connectors
The battery is a sealed, lead-acid battery and must be recycled or
disposed of properly according to national or local regulations.