Use the New Patient tab to perform an Auto ECG test
CAUTION Patient data is not saved until the ECG test is completed.
Note The ECG configuration settings can be changed in the Settings tab. The
following settings may appear differently if the default settings have been
modified.
Note Set the default patient entry tab to New patient in the Advanced settings.
1. Touch (Auto ECG). The New patient tab appears.
Note In a connected environment, with the default patient entry
tab set to Worklist (in the Advanced settings), the worklist is
downloaded from the Worklist server workstation and the
Worklist tab appears. Touch the New patient tab to proceed
with the New patient workflow.
2. Enter the following patient information as desired:
Patient ID. Touch OK.
Birth date. Touch OK.
Gender. Touch OK.
Last name. Touch OK.
First name. Touch OK.
Middle Initial. Touch OK.
Note If the patient has a pacemaker touch Pacemaker present.
3. Touch (Next).
4. Enter the following patient information as desired:
Race
Height. Touch OK.
Weight. Touch OK.
Physician. Touch OK.
Comments. Touch OK.
5. Attach the leads to the patient.
6. (Optional) Adjust the waveforms, using the buttons to cycle through the following
options:
leads displayed
ECG preview format
gain (size)
speed
filters
26 ECG tests CP 150 12-lead resting electrocardiograph